Home

CT Scanner Setup Examples

Scanner Setup is a tool to optimize the tradeoff between acquisition speed and image quality by adjusting source KV, filter choice, specimen size, and detector stopping power.

Example 1: Simulation of micro CT of carbonate core plug using a typical laboratory scanner

Specimen is relatively homogeneous. Source Max KV 160, Detector CsI screen 100μm thick
Intial Setup: 160 KV, 0.01cm Cu filter

Adjustment: Increase filter thickness to 0.1cm to reduce beam hardening.

Adjustment: 160 KV, 0.10cm Cu filter

Exercises:


Example 2: Simulation of an 3cm aluminum extrudate with internal iron and brass pins

Specimen is heterogeneous and highly structured. The brass pins are rectangular and about 0.4cm in the long direction. When scanned at conditions same as example 1, the high opacity of the brass pins causes strong beam-hardening streak artifacts.
Extrudate 160 KV, 0.10cm Cu filter
The noiseless simulation showed a peak opacity τ ~8. The artifact on a real scanner would most likely been much more severe.
The Table below shows a series of steps to reduce both τ and beam-hardening while increasing photon use. Note the detector change from thin CsI to thick cadmium tungstate
Start(top row) to Finish(bottom row) Setup Changes for Improved Scan
The plot below shows the spectral distributions at the final configuration. The detector's high-energy response is improved. Most of the low energy photons have been removed by the filter.
Spectral Distributions for Improved Scan
Extrudate 250 KV, 1.0cm Sn filter, CadTungstate detector
No post-scan beam hardening correction was applied.

1. My Excel Beam Hardening Add-in uses the Solver to maximize Photon Use% while constraining sample tau and beam hardening.

Home